Soft X-ray response of a CCD with a grating spectrometer

Citation
M. Shouho et al., Soft X-ray response of a CCD with a grating spectrometer, NUCL INST A, 436(1-2), 1999, pp. 85-90
Citations number
10
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
436
Issue
1-2
Year of publication
1999
Pages
85 - 90
Database
ISI
SICI code
0168-9002(19991021)436:1-2<85:SXROAC>2.0.ZU;2-V
Abstract
We calibrate the X-ray imaging spectrometers, which are CCD cameras install ed on the ASTRO-E satellite, by using dispersed continuous soft X-rays from a grating spectrometer. We obtained the signal-pulse height and energy-res olution as a function of X-ray energies continuously. However, the wings of the line spread function of the grating distorts the center of the signal- pulse height derived by a simple analysis. An estimation of this distortion is presented. We also describe two methods of extracting the pure signal-p ulse-height distribution from the data using the spectrometer. A brief desc ription of the low-energy tail is presented. (C) 1999 Elsevier Science B.V. All rights reserved.