Liquid helium transfer using a mass-flow controller

Citation
Y. Fujii et al., Liquid helium transfer using a mass-flow controller, REV SCI INS, 70(10), 1999, pp. 4110-4113
Citations number
6
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
REVIEW OF SCIENTIFIC INSTRUMENTS
ISSN journal
00346748 → ACNP
Volume
70
Issue
10
Year of publication
1999
Pages
4110 - 4113
Database
ISI
SICI code
0034-6748(199910)70:10<4110:LHTUAM>2.0.ZU;2-3
Abstract
An automated liquid helium transfer method using a mass-flow controller, by means of which an empty cryostat at room temperature can be filled with li quid helium without thermal shock, has been developed. Before liquid helium collection is initiated, mass flow is controlled to be constant. After the start of liquid helium collection, mass flow is controlled so that the dif ferential pressure of the empty cryostat against the storage Dewar containi ng the liquid helium is of the proper value to overcome the height differen ce due to the transfer tube. A mass-flow controller, a vacuum pump, a diffe rential pressure sensor, a liquid helium level sensor and a computer with s uitable software are required to replace the conventional manual transfer b y this method. (C) 1999 American Institute of Physics. [S0034- 6748(99)0331 0-9].