The calibration of several ions (Cl-, Br-, F- and OH-) measured with an ion
selective electrodes (ISE) array has been carried out in the presence of i
nterferents using an experimental design and multivariate calibration metho
ds. Partial least squares regression and principal component regression do
not seem to improve the test set prediction compared to multivariate linear
regression. In the case of very slight or no interference on the ISE, each
ion can be determined using the corresponding ISE and univariate calibrati
on methods, but the use of multivariate methods does not lead to worse resu
lts. (C) 1999 Published by Elsevier Science B.V. All rights reserved.