T. Masadome et al., Flow injection determination of bromide ion in a developer using bromide ion-selective electrode detector, TALANTA, 50(3), 1999, pp. 595-600
A potentiometric flow injection determination method for bromide ion in a d
eveloper was proposed, by utilizing a flow-through type bromide ion-selecti
ve electrode detector. The sensing membrane of the electrode was Ag2S-AgBr
membrane. The response of the electrode detector as a peak-shape signal was
obtained for injected bromide ion in a developer. A linear relationship wa
s found to exist between peak height and the concentration of the bromide i
on in a developer in a concentration range from 1.0 x 10(-3) to 1.0 x 10(-2
) mol l(-1). The relative standard deviation for 10 injections of a 6 x 10-
3 mol l(-1) bromide ion in a developer was 1.3% and the sampling rate was c
a 17-20 samples h(-1). The present method was free fr om the interference o
f an organic reducing reagent, an organic substance in a developer sample s
olution for the determination of bromide ion in a developer. (C) 1999 Publi
shed by Elsevier Science B.V. All rights reserved.