Da. Jaffray et al., A radiographic and tomographic imaging system integrated into a medical linear accelerator for localization of bone and soft-tissue targets, INT J RAD O, 45(3), 1999, pp. 773-789
Citations number
37
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Onconogenesis & Cancer Research
Journal title
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS
Purpose:Dose escalation in conformal radiation therapy requires accurate fi
eld placement. Electronic portal imaging devices are used to verify field p
lacement but are limited by the low subject contrast of bony anatomy at meg
avoltage (MV) energies, the large imaging dose, and the small size of the r
adiation fields. In this article, we describe the in-house modification of
a medical linear accelerator to provide radiographic and tomographic locali
zation of bone and soft-tissue targets in the reference frame of the accele
rator. This system separates the verification of beam delivery (machine set
tings, field shaping) from patient and target localization.
Materials and Methods: A kilovoltage (kV) x-ray source is mounted on the dr
um assembly of an Elekta SL-20 medical linear accelerator, maintaining the
same isocenter as the treatment beam with the central axis at 90 degrees to
the treatment beam axis. The x-ray tube is powered by a high-frequency gen
erator and can be retracted to the drum-face. Two CCD-based fluoroscopic im
aging systems are mounted on the accelerator to collect MV and kV radiograp
hic images. The system is also capable of cone-beam tomographic imaging at
both MV and kV energies. The gain stages of the two imaging systems have be
en modeled to assess imaging performance. The contrast-resolution of the kV
and MV systems was measured using a contrast-detail (C-D) phantom, The dos
imetric advantage of using the kV imaging system over the MV system for the
detection of bone-like objects is quantified for a specific imaging geomet
ry using a C-D phantom. Accurate guidance of the treatment beam requires re
gistration of the imaging and treatment coordinate systems. The mechanical
characteristics of the treatment and imaging gantries are examined to deter
mine a localizing precision assuming an unambiguous object. MV and kV radio
graphs of patients receiving radiation therapy are acquired to demonstrate
the radiographic performance of the system. The tomographic performance is
demonstrated on phantoms using both the MV and the kV imaging system, and t
he visibility of soft-tissue targets is assessed.
Results and Discussion: Characterization of the gains in the two systems de
monstrates that the MV system is x-ray quantum noise-limited at very low sp
atial frequencies; this is not the case for the kV system. The estimates of
gain used in the model are validated by measurements of the total gain in
each system. Contrast-detail measurements demonstrate that the MV system is
capable of detecting subject contrasts of less than 0.1% (at 6 and 18 MV).
A comparison of the kV and MV contrast-detail performance indicates that e
quivalent bony object detection can be achieved with the kV system at signi
ficantly lower doses (factors of 40 and 90 lower than for 6 and 18 MV, resp
ectively). The tomographic performance of the system is promising; soft-tis
sue visibility is demonstrated at relatively low imaging doses (3 cGy) usin
g four laboratory rats.
Conclusions: We have integrated a kV radiographic and tomographic imaging s
ystem with a medical linear accelerator to allow localization of bone and s
oft-tissue structures in the reference frame of the accelerator. Modeling a
nd experiments have demonstrated the feasibility of acquiring high-quality
radiographic and tomographic images at acceptable imaging doses. Full integ
ration of the kV and MV imaging systems with the treatment machine will all
ow on-line radiographic and tomographic guidance of field placement. (C) 19
99 Elsevier Science Inc.