A new structure transcutaneous transformer for artificial heart system

Citation
L. Zhao et al., A new structure transcutaneous transformer for artificial heart system, IEEE MAGNET, 35(5), 1999, pp. 3550-3552
Citations number
7
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science
Journal title
IEEE TRANSACTIONS ON MAGNETICS
ISSN journal
00189464 → ACNP
Volume
35
Issue
5
Year of publication
1999
Part
2
Pages
3550 - 3552
Database
ISI
SICI code
0018-9464(199909)35:5<3550:ANSTTF>2.0.ZU;2-X
Abstract
A new geometrical design of transcutaneous transformer for artificial heart system is presented in this paper. The primary and secondary cores are mad e of ferrite and amorphous magnetic materials respectively. The experimenta l results indicate that it is possible to transmit an electrical power of 2 0 W which is sufficient to drive the heart pump. With a gap of Smm between the coils, its transmission efficiency with the inverter can be up to 90% w ith a switching frequency 102.5kHz. Some characteristics of the proposed tr ansformer are also shown and discussed in this paper.