Today most surgical treatment of spinal deformations is concentrated on inv
asive techniques with long operation times and major effects on the patient
's mobility. The proposed minimally invasive technique using laser light fo
r tissue ablation offers a possibility of gentle scoliosis treatment. Four
different laser systems including argon, Nd:YAG (e-switched), Nd:YAG (CW) a
nd Ho:YAG were compared with respect to thermal damage to adjacent tissue,
ablation rates, efficiency and laser handling. For in vitro investigation,
fresh lamb spine was used, Comparison showed that the Ho:YAG laser is the m
ost appropriate laser for the given goal, however, heating of adjacent tiss
ue had to be minimized by intermediate laser action and irrigation.