Optimization of gray/white matter contrast with fast inversion recovery for myelin suppression: A comparison of fast spin-echo and echo-planar MR imaging sequences
Lj. Wolansky et al., Optimization of gray/white matter contrast with fast inversion recovery for myelin suppression: A comparison of fast spin-echo and echo-planar MR imaging sequences, AM J NEUROR, 20(9), 1999, pp. 1653-1657
Citations number
14
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Neurosciences & Behavoir
We compared two MR imaging sequences, fast inversion recovery for myelin su
ppression (FIRMS) and echo-planar FIRMS (EP-FIRMS), for depicting gray/whit
e matter contrast. In 18 patients, the frequency bandwidth (BW) was optimiz
ed for each sequence; in nine patients, the BW was held constant. In the BW
-optimized group, the mean contrast-to-noise ratio (C/N) was three times hi
gher with the FIRMS sequence. In the BW-constant group, the mean C/N was 27
% higher with the FP-FIRMS sequence; however, geometric distortion degraded
the EP-FIRMS images excessively. For optimal gray/white contrast, FIRMS ap
pears to be the superior pulse sequence.