Mp. Flament et al., Influence of the technological parameters of ultrasonic nebulisation on the nebulisation quality of alpha 1 protease inhibitor (alpha 1I), INT J PHARM, 189(2), 1999, pp. 197-204
The principle of an ultrasonic nebuliser is based on the vibrations of a pi
ezo-electric crystal driven by an alternating electrical field. These perio
dical vibrations are characterised by their frequency, their amplitude and
their intensity which corresponds to the energy transmitted per surface uni
t. When the vibration intensity is sufficient, cavitation appears and gener
ates droplets. Ventilation enables an airflow to cross the nebuliser and to
expulse the aerosol droplets. For a given nebuliser, the vibration frequen
cy of the piezo-electric crystal is fixed and is often in the range of 1-2.
5 MHz. In most cases, an adjustment in vibration intensity is possible by m
odifying vibration amplitude. The ventilation level is adjustable. The infl
uence of these two parameters on the efficiency of ultrasonic nebulisation
is studied. The study was carried out with a protein solution that had to b
e administered into the lungs. The solution used presented a viscosity of 1
.25 mPa and a surface tension of 53 mN/m. The integrity of the protein was
checked which was submitted to different vibration conditions. Nebulisation
efficiency was evaluated by determining droplet size, the percentage of dr
ug nebulised and nebulisation time. An increase in vibration intensity does
not modify the size of droplets emitted, but decreases nebulisation time a
nd raises the quantity of protein nebulised, thus improving performance. On
the other hand, an increase in ventilation increases the size of droplets
emitted, decreases nebulisation time and the quantity of protein nebulised
because more drug is lost on the walls of the nebuliser. High intensity ass
ociated with low ventilation favours drug delivery deep into the lungs. (C)
1999 Elsevier Science B.V. All rights reserved.