Coronary dissection due to overdilatation of an elastic membrane of the multi-link stent delivery system caused by balloon rupture

Citation
N. Kuroda et al., Coronary dissection due to overdilatation of an elastic membrane of the multi-link stent delivery system caused by balloon rupture, JPN CIRC J, 63(11), 1999, pp. 912-913
Citations number
8
Categorie Soggetti
Cardiovascular & Respiratory Systems","Cardiovascular & Hematology Research
Journal title
JAPANESE CIRCULATION JOURNAL-ENGLISH EDITION
ISSN journal
00471828 → ACNP
Volume
63
Issue
11
Year of publication
1999
Pages
912 - 913
Database
ISI
SICI code
0047-1828(199911)63:11<912:CDDTOO>2.0.ZU;2-X
Abstract
A 65-year-old male with unstable angina underwent coronary angiography, whi ch revealed a significant stenotic lesion in the right coronary artery. Thi s narrowing was subsequently treated with the Multi-Link stent. During the balloon inflation associated with stent deployment, balloon rupture occurre d and resulted in overdilatation of an elastic membrane in the stent delive ry system. This, in turn, resulted in coronary dissection? which required t reatment with further stenting.