Jr. Castillo et al., Semiquantitative simultaneous determination of metals in olive oil using direct emulsion nebulization, J ANAL ATOM, 14(9), 1999, pp. 1515-1518
A simple method for the semiquantitative determination of several metals in
olive oil by ICP-MS is described. The formation of olive oil emulsions, as
a simplified sample preparation procedure is demonstrated. Optimization of
the chemical conditions for the formation and stabilization of olive oil e
mulsions is reported: pH, emulsifiers, stirring (mechanical and ultrasonic
mode) time, reagents and sample volumes. The semiquantitative determination
of Ba, Cd, Co, Cr, Cu, Mn, Ni, Tl, Pb, U, V and Zn was carried out by ICP-
MS with direct introduction of the emulsions into the plasma after optimiza
tion of the instrumental parameters (power, pump flow, nebulizer gas flow,
use of auxiliary oxygen flow). Results from spike and recovery experiments
at levels of 100 and 200 ng mL(-1) are 93-115% and detection limits based o
n 3 times standard deviation of the blank (10 replicates) are 0.25-50 ng g(
-1) in the olive oil.