Vp. Vassilikos et al., Percutaneous extraction of transvenous defibrillator leads using the VascoExtor pacing lead removal system, J INTERV C, 3(3), 1999, pp. 247-251
Citations number
8
Categorie Soggetti
Cardiovascular & Respiratory Systems
Journal title
JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY
In the implantable cardioverter defibrillator era the necessity for lead re
moval is not negligible. A specially designed extraction lead system for pe
rcutaneous removal of such leads is lacking, in contrast to the existing pa
cing lead extraction systems. We report the successful percutaneous extract
ion of four implantable cardioverter defibrillator leads in three patients
because of lead malfunction using a novel pacemaker lead extraction system,
the VascoExtor (VascoMed) system. Three leads were successfully removed in
two patients using traction with special locking stylets from the superior
approach. One lead was removed using the system's additional extraction to
ols through the femoral approach. There were no complications. This prelimi
nary experience shows that the VascoExtor (VascoMed) pacemaker lead extract
ion system can also be used in implantable cardioverter defibrillator lead
extraction safely and effectively. In addition to the locking stylets, adju
nct percutaneous extraction tools may be needed in some cases.