Effects of transducer frequency and output power on the ultrasonographic contrast produced by Optison using fundamental and harmonic imaging techniques

Citation
Jh. Wible et al., Effects of transducer frequency and output power on the ultrasonographic contrast produced by Optison using fundamental and harmonic imaging techniques, J ULTR MED, 18(11), 1999, pp. 753-762
Citations number
18
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging
Journal title
JOURNAL OF ULTRASOUND IN MEDICINE
ISSN journal
02784297 → ACNP
Volume
18
Issue
11
Year of publication
1999
Pages
753 - 762
Database
ISI
SICI code
0278-4297(199911)18:11<753:EOTFAO>2.0.ZU;2-G
Abstract
This in vivo study demonstrated the effects of transducer frequency and out put power on the sonographic contrast produced by Optison during continuous ultrasonographic imaging in both fundamental and harmonic modes. During fu ndamental imaging, Optison produced greater contrast intensity while imagin g at higher transducer frequencies. However, in the harmonic mode, lower im aging frequencies caused Optison to produce more intense myocardial enhance ment. Regardless of imaging mode, increases in transducer output power caus ed decreases in both contrast intensity and contrast duration produced by O ptison. Proper selection of transducer frequency and output power will help optimize the quality and duration of contrast produced by Optison during c ontinuous ultrasonographic imaging.