Backround: Although cataract surgery is highly developed today, there are s
till problems such as secondary cataract. In polishing the posterior capsul
e lens, epithelial cells often remain, causing secondary cataracts. Additio
nally, there are the problems of tissue hearing and endothelial cell loss d
uring phacoemulsification by ultrasound. This could be another field for im
provement of cataract surgery by using the waterjet.
Methods: After removing the cornea of freshly enucleated porcine bulbs, we
used the water jet from 4 to 12 bar. The hit angel on the capsule varied be
tween 45 degrees and 90 degrees.
Results: Rupture of the posterior capsule occurred at a mean of 8.5 bar usi
ng the jet at 45 degrees and a mean of 8.6 bar using the jet at 90 degrees.
Conclusion: The waterjet pressure should not be over 4 bar during polishing
of the posterior capsule.