A new multiparameter sensor that combines electrochemical and fibre-optic t
echnology was used for continuous in-vivo investigation of pH, carbon dioxi
de partial pressure (PCO2), oxygen partial pressure (PO2), bicarbonate conc
entration (HCO3-), base excess, and oxygen saturation (O(2)Sat) early in hu
man pregnancy. The sensor was inserted into the amniotic cavity and the pla
cental bed of 16 pregnancies at 10-15 weeks gestation, before termination u
nder general anaesthesia, Amniotic fluid and retroplacental blood from the
same site were also aspirated and analysed by means of cartridges and a por
table blood gas analyser. Eleven series of measurements were obtained. The
variation in measurements over the 5 min of monitoring was less than or equ
al to 10% for all parameters. The sensor was damaged during insertion into
the amniotic cavity in one case and in the placental bed in four cases. Mea
surements of PO2 in both the amniotic cavity and the placental bed and of p
H in the placental bed were higher using the cartridges than in vivo. The r
esults indicated that in-vivo monitoring of fetoplacental gas and acid-base
with a sensor is stable and accurate. Such technology will be helpful in i
mproving our understanding of the fetoplacental metabolism in normal and co
mplicated pregnancies.