Hj. Pfaeffle et al., A new methodology to measure load transfer through the forearm using multiple universal force sensors, J BIOMECHAN, 32(12), 1999, pp. 1331-1335
Previous approaches to measuring forces in the forearm have made the assump
tion that forces acting in the radius and ulna are uniaxial near the wrist
and elbow. To accurately describe forces in the forearm and the forces in t
he interosseous ligament, we have developed a new methodology to quantitati
vely determine the 3-D force vectors acting in forearm structures when a co
mpressive load is applied to the hand. A materials testing machine equipped
with a six degree-of-freedom universal force-moment sensor (UFS) was emplo
yed to apply a uniaxial compressive force to cadaveric forearms gripped at
the hand and humerus. Miniature UFSs were implanted into the distal radius
and proximal ulna to measure force vectors there. A 3-D digitizing device w
as used to measure transformations between UFS coordinate systems, utilized
for calculating the force vectors in the distal ulna, proximal radius, and
the interosseous ligament (IOL). This method was found to be repeatable to
within 3 N, and accurate to within 2 N for force magnitudes. Computer mode
ls of the forearm, generated from CT scans, were used to visualize the forc
e vectors in 3-D. Application of this methodology to eight forearm specimen
s showed that the radius carries most of the load at the wrist while force
in the IOL relieves load acting in the radius at the mid-forearm. For a 136
N applied hand force, the force in the IOL was 36 +/- 21 N. Advantages of
this methodology include the determination of 3-D force vectors, especially
those in the IOL: as well as computer generated 3-D visualization of resul
ts. (C) 1999 Elsevier Science Ltd. All rights reserved.