The response and calibration of thin Si Delta E detectors

Citation
Yw. Zhang et al., The response and calibration of thin Si Delta E detectors, NUCL INST B, 159(1-2), 1999, pp. 101-110
Citations number
47
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS
ISSN journal
0168583X → ACNP
Volume
159
Issue
1-2
Year of publication
1999
Pages
101 - 110
Database
ISI
SICI code
0168-583X(199910)159:1-2<101:TRACOT>2.0.ZU;2-8
Abstract
The response of thin (similar to 7 mu m) self-supporting Si p-i-n Gamma E d etectors for light recoils (Z less than or equal to 8) has been characteris ed using a conventional Elastic Recoil Detection (ERD) set-up. In this stud y, the energy loss of each recoil in the Delta E detector was measured by d etermining the incident energy from the time of flight and the energy after traversing the Delta E detector using a thick (280 mu m) Si p-i-n diode E detector. 40 MeV Cl-35(7+) ions were used to produce recoils of H, Li, Be, B, C, N and O. Correlation of the energy loss in the Delta E detector with the amplitude of the Delta E signals was used to assign calibration constan ts. Analysis of the variance of the signals from the trend line revealed th at the standard deviation of the data perpendicular to the trend line which is not contributed to by energy straggling and multiple scattering, increa ses with increasing recoil atomic number. (C) 1999 Elsevier Science B.V. Al l rights reserved.