S. Poyard et al., Association of a poly(4-vinylpyridine-co-styrene) membrane with an inorganic/organic mixed matrix for the optimization of glucose biosensors, SENS ACTU-B, 58(1-3), 1999, pp. 380-383
The association of an inorganic/organic composite matrix with a permselecti
ve poly(4-vinylpyridine-co-styrene) membrane allows to obtain a selective g
lucose biosensor, allowing a sensitivity of 30 mA M-1 cm(-2), for a linear
dynamic range from 10(-5) M to 1.5 mM. Moreover, the combined use of such p
ermselective membranes with a bienzymatic system, glucose oxidase/polypheno
l oxidase, has demonstrated that the effect of electroactive interferent sp
ecies such as ascorbate, mate and p-acetaminophen, can be entirely removed.
(C) 1999 Elsevier Science S.A. All rights reserved.