Development of a dissolved carbon dioxide sensor based on a coulometric titration

Citation
K. Wiegran et al., Development of a dissolved carbon dioxide sensor based on a coulometric titration, SENS ACTU-B, 57(1-3), 1999, pp. 120-124
Citations number
13
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
SENSORS AND ACTUATORS B-CHEMICAL
ISSN journal
09254005 → ACNP
Volume
57
Issue
1-3
Year of publication
1999
Pages
120 - 124
Database
ISI
SICI code
0925-4005(19990907)57:1-3<120:DOADCD>2.0.ZU;2-A
Abstract
This paper presents a sensor for the determination of dissolved carbon diox ide in water and seawater, respectively. The gas diffuses through a hydroph obic gas-permeable membrane and reacts with an alkaline electrolyte causing a ptr-value shift which is detected by an iridium oxide pH-electrode. The duration of the pH-shift within a specified pH-interval is measured and thi s so-called measuring period is correlated with the partial pressure of dis solved carbon dioxide in the sample. To start a new cycle, the initial pH-v alue is reached again by generating hydroxide ions electrochemically. The s ensor allows determination of dissolved carbon dioxide in water in the rang e from 30 to 180 ppm. (C) 1999 Elsevier Science S.A. All rights reserved.