Time domain self-reciprocity calibration of ultrasonic transducer

Citation
Mx. Deng et al., Time domain self-reciprocity calibration of ultrasonic transducer, ACT PHY C E, 8, 1999, pp. S48-S52
Citations number
10
Categorie Soggetti
Physics
Journal title
ACTA PHYSICA SINICA-OVERSEAS EDITION
ISSN journal
10003290 → ACNP
Volume
8
Year of publication
1999
Supplement
S
Pages
S48 - S52
Database
ISI
SICI code
1000-3290(199908)8:<S48:TDSCOU>2.0.ZU;2-W
Abstract
In this article, the method of time domain self-reciprocity calibration is given. A time domain echo voltage can be obtained once a time domain excita tion current flows through an ultrasonic transducer (UT). The current. is a ccurately measured by a special current clamp. The UT sensitivities can be obtained by performing Fourier transform of the time domain excitation curr ent and the time domain echo voltage. The calibration method used in this a rticle possesses such advantages as high calibration speed, wide frequency band and simple operation procedure. Moreover, there is an obvious merit th at the UT phase-frequency curve could be obtained. By comparing the results of time domain self-reciprocity calibration with those obtained by using f requency domain method, it is found that two results well agree with each o ther.