Clinical implementation of a computer controlled milling machine for compensating filter production

Citation
M. Partridge et al., Clinical implementation of a computer controlled milling machine for compensating filter production, BR J RADIOL, 72(863), 1999, pp. 1099-1103
Citations number
8
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
BRITISH JOURNAL OF RADIOLOGY
ISSN journal
00071285 → ACNP
Volume
72
Issue
863
Year of publication
1999
Pages
1099 - 1103
Database
ISI
SICI code
Abstract
The procedures required for the clinical implementation of a computer contr olled milling machine for producing compensators for breast radiotherapy ar e described. Moulds are cut in a rigid polymer foam block and filled with s tainless-steel granulate. Quality assurance procedures are described for en suring that the compensators produced are consistent and accurate. Relative and absolute dosimetric measurements are presented showing that the compen sators are accurate to better than 1% and demonstrate the technique to be c linically acceptable.