M. Partridge et al., Clinical implementation of a computer controlled milling machine for compensating filter production, BR J RADIOL, 72(863), 1999, pp. 1099-1103
Citations number
8
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
The procedures required for the clinical implementation of a computer contr
olled milling machine for producing compensators for breast radiotherapy ar
e described. Moulds are cut in a rigid polymer foam block and filled with s
tainless-steel granulate. Quality assurance procedures are described for en
suring that the compensators produced are consistent and accurate. Relative
and absolute dosimetric measurements are presented showing that the compen
sators are accurate to better than 1% and demonstrate the technique to be c
linically acceptable.