Methohexital (MS) determination is based on the formation of insoluble merc
ury salt on a hanging mercury drop electrode after preaccumulation by adsor
ption. This property was exploited in developing a highly sensitive strippi
ng voltammetric procedure for the determination of the drug. The anodic cur
rent of adsorbed compound is measured by linear sweep anodic stripping volt
ammetry (LSASV), preceded by a period of preconcentration. The effect of va
rious parameters such as supporting electrolyte composition, pH, initial po
tential, scan rate, accumulation time and ionic strength are discussed to c
haracterize the interfacial and redox behavior. The detection limit was fou
nd to be 2 x 10(-7) M (56.8 ppb) with 180-s accumulation time. The interfer
ence of some amino acids, ascorbic acid and some metal ions was investigate
d. The application of this method was tested in the determination of methoh
exital in spiked urine samples. The precision of the method is satisfactory
with a relative standard deviation of 2.5%. (C) 1999 Elsevier Science B.V.