A diagnostic beamline has been operated in the Pohang Light Source (PLS) st
orage ring for the diagnostics of electron and photon beam properties. It c
onsists of two 1:1 imaging systems: a visible-light imaging system and a so
ft X-ray imaging system. We have measured the transverse and the longitudin
al structures of beams by using a streak camera to obtain a visible image.
Accurate transverse beam sizes have been measured to be 186 mu m horizontal
ly and 43.1 mu m vertically by using soft X-ray images with minimum diffrac
tion errors. The col responding emittances are 11.7 nm-rad horizontally and
0.59 nm-rad vertically. By comparing the measured data with the design val
ues, we confirmed that the PLS storage ring has reached its desgn performan
ce within an error of 3.3% in the transverse direction.