Portal dosimetry using x-ray film: An experimental and computational study
Citation
Yp. Zhu et Ji. Yeo, Portal dosimetry using x-ray film: An experimental and computational study, MED PHYS, 26(11), 1999, pp. 2403-2409
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
MEDICAL PHYSICS
SICI code
0094-2405(199911)26:11<2403:PDUXFA>2.0.ZU;2-P
Abstract
To evaluate the accuracy and precision of relative portal dosimetry using x
-ray film, we compared the radiation doses measured by x-ray film and by an
ion chamber at various portal planes that were 0 to 50 cm behind an 18 cm
thick phantom. In addition, we calculated photon spectra at the measurement
planes by using the Monte Carlo particle transport technique. The experime
nts showed that the film usually measured relative doses to within 5% of th
e measurements by the ion chamber and that the errors were associated with
the changes in the spectra of photon energy fluence at portal planes. The r
elative magnitude of low-energy fluence in the photon fluence spectrum and
its variance across the portal plane caused the film response to differ fro
m the ion-chamber response especially with lead screen on top of film in th
e film cassette. Relative film dosimetry may be improved to accuracy of bet
ter than 2% by using solid water or other tissue equivalent cassettes. (C)
1999 American Association of Physicists in Medicine. [S0094-2405(99)00411-3
].