Assessment of the strength of individual Ir-192 seeds in ribbons

Citation
B. Thomadsen et al., Assessment of the strength of individual Ir-192 seeds in ribbons, MED PHYS, 26(11), 1999, pp. 2471-2475
Citations number
3
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
MEDICAL PHYSICS
ISSN journal
00942405 → ACNP
Volume
26
Issue
11
Year of publication
1999
Pages
2471 - 2475
Database
ISI
SICI code
0094-2405(199911)26:11<2471:AOTSOI>2.0.ZU;2-8
Abstract
Assessing the strength of individual seed-type sources in ribbon assembles remains a challenge in brachytherapy quality assurance. Geometries to measu re a single source in the ribbon usually fail because of low signals if usi ng very thick shielding to block the radiation from the other sources, or c ontributions from ail the other sources if they are not shielded well. A no rmal well-type chamber with partial lead shielding forming a small slot pro vides a differential response along the chamber axis that, through a deconv olution/simultaneous-equations technique, sorts the contributions from each source, allowing the derivation of each source's strength. (C) 1999 Americ an Association of Physicists in Medicine. [S0094-2405(99)03311-8].