Micro channel plate gains for Ta10+-Ta44+ ions, measured in the energy range from 3.7 keV/q up to 150.7 keV/q

Citation
W. Mroz et al., Micro channel plate gains for Ta10+-Ta44+ ions, measured in the energy range from 3.7 keV/q up to 150.7 keV/q, NUCL INST A, 437(2-3), 1999, pp. 335-345
Citations number
24
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT
ISSN journal
01689002 → ACNP
Volume
437
Issue
2-3
Year of publication
1999
Pages
335 - 345
Database
ISI
SICI code
0168-9002(19991121)437:2-3<335:MCPGFT>2.0.ZU;2-R
Abstract
The procedure of calibration of a detector assembly consisting of a micro c hannel plate detector operated in a Chevron configuration is described. The analog particle gains of Ta ions with charge states ranging from q = 10(+) to 44(+) and ion impact energies from 3.2 to 153.2 keV/q have been measure d. For the investigated range of ion impact energies we have stated that fo r tantalum charge states of 10(+)-20(+) the secondary ion-electron emission coefficient is dominated by the kinetic emission of electrons. Potential e ffects being visible for the charge states above 20(+). (C) 1999 Elsevier S cience B.V. All rights reserved.