Fb. Mohamed et al., A-simple method to improve image nonuniformity of brain MR images at the edges of a head coil, J COMPUT AS, 23(6), 1999, pp. 1008-1012
Citations number
14
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
One of the major sources of image nonuniformity in the high field MR scanne
rs is the radiofrequency (RF) coil inhomogeneity. It degrades conspicuity c
if: lesion(s) in the MR images of the brain and surrounding tissues and red
uces accuracy of image postprocessing particularly at the edges of the coil
. In this investigation, we have devised and tested a simple method to corr
ect for nonuniformity of MR images of the brain at the edges of the RF head
coil. Initially, a cylindrical oil phantom, which fit exactly in the head
coil, was scanned on a 1.5 T imager. Then, a correction algorithm identifie
d a reference pixel value in the phantom at the most homogeneous region of
the RF coil. Next, every pixel inside the phantom was normalized relative t
o this reference value. The resulting set of coefficients "correction matri
ces'' was obtained for different types of MR contrast agent. Finally, brain
MR images of normal subjects and multiple sclerosis patients were acquired
and processed by the corresponding correction matrices obtained with diffe
rent pulse sequences. Application of correction matrices to brain MR images
showed a gain in pixel intensity particularly in the slices at the edge of
the coil.