Bx. Li et al., Flow injection chemiluminescence determination of isoniazid using on-line electrogenerated manganese(III) as oxidant, MICROCHEM J, 63(3), 1999, pp. 374-380
A novel how injection chemiluminescence (CL) system for the determination o
f isoniazid has been proposed. It is based on the direct CL reaction of iso
niazid and Mn(III) in sulfuric acid medium. The unstable Mn(III) was on-lin
e electrogenerated by constant current electrolysis. The CL emission intens
ity was Linear with isoniazid concentration in the range 0.1-10 mu g/mL; th
e detection limit was 3.2 x 10(-2) mu g/mL. The whole process could be comp
leted in 1 min with a relative standard deviation of less than 5%. The prop
osed method is suitable for automatic and continuous analysis and has been
applied successfully to the analysis of isoniazid in pharmaceutical prepara
tion. (C) 1999 Academic Press.