Jc. Wolff et al., New ultraviolet signal actuated switching valve for the measurement of lowlevel impurities by liquid chromatography/mass spectrometry, RAP C MASS, 13(23), 1999, pp. 2376-2381
A new ultraviolet (UV) signal actuated switching valve for diverting the ma
in matrix compound to waste, preventing it entering the ion source of the m
ass spectrometer is described. Sensitivity for trace impurities eluting aft
er the drug substance cimetidine or related compounds could be enhanced by
a factor of 4-5. The increase in sensitivity was dependent on the type of i
on source which interfaced the mass spectrometer, The benefit of the switch
ing valve was greater with a line of sight type source than with an orthogo
nal one. The detection limit for a trace compound in a matrix compound was
improved by up to a factor of 10 with the line of sight type source but onl
y by a factor of 5 with the orthogonal source. Copyright (C) 1999 John Whey
& Sons, Ltd.