Radioluminescence detector for the flow injection determination of phosphorus as vanadomolybdophosphoric acid

Citation
Am. Leach et al., Radioluminescence detector for the flow injection determination of phosphorus as vanadomolybdophosphoric acid, ANALYT CHIM, 402(1-2), 1999, pp. 267-274
Citations number
23
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICA CHIMICA ACTA
ISSN journal
00032670 → ACNP
Volume
402
Issue
1-2
Year of publication
1999
Pages
267 - 274
Database
ISI
SICI code
0003-2670(199912)402:1-2<267:RDFTFI>2.0.ZU;2-A
Abstract
A radioluminescent light source has been used in a detector for the how inj ection (FI) determination of phosphorus by means of the vanadomolybdophosph oric (VMP) acid spectrophotometric method. A comparison with previous studi es in which lamp-based instrumentation was utilized proves that radiolumine scence (RL) is a viable alternative source. RL enables the design of compac t, inexpensive light sources for spectroscopic studies. Through the selecti on of a radioisotope and scintillation medium, the spectral and temporal ch aracteristics of the source can be chosen independently. In this study, a b roadband radioluminescent source provided detection limits similar to those of conventional lamp-based detectors while increasing the dynamic range of the method. Possible interferences caused by the use of a broadband light source have been explored as a function of the spectral region employed. (C ) 1999 Elsevier Science B.V. All rights reserved.