Isodispersion curves in flow injection analysis

Citation
Di. Vaireanu et al., Isodispersion curves in flow injection analysis, ANAL COMMUN, 36(11-12), 1999, pp. 407-408
Citations number
4
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
ANALYTICAL COMMUNICATIONS
ISSN journal
13597337 → ACNP
Volume
36
Issue
11-12
Year of publication
1999
Pages
407 - 408
Database
ISI
SICI code
1359-7337(1999)36:11-12<407:ICIFIA>2.0.ZU;2-G
Abstract
This paper describes experimental equations that relate the dispersion coef ficient to the geometry and hydrodynamic parameters of a flow injection sys tem. Data are presented as isodispersion curves (similar to isothermal, iso bar or isochore representations). The results present a correlated overview , which makes theoretical evaluation of the dispersion coefficient simple. Geometric and operational parameters for a flow injection system can be est ablished when the dispersion is to be maintained within certain limits. The parameters addressed in this study are: tube length (100-1200 mm); tube di ameter (0.3-0.8 mm); injected volume (93 mu l); and volumetric flow rate (0 .75-2.0 ml min(-1)).