The magnitude of serum thyroxine (T4) binding capacity (sBC) dependent bias
in the AXSYM free thyroxine (FT4) assay was assessed using two recently de
scribed tests. One of the tests uses a direct equilibrium dialysis (ED) FT4
assay as the reference method. The results obtained with the AXSYM method
were compared with those obtained by the ED FT4 method in patient sera havi
ng a wide range of sBC The other test involves comparison of the FT4 result
s obtained following dilution of sera by an inert buffer, to theoretically
derived FT4 results. As serum dilution causes a predictable decrease in sBC
, the demonstration of a negative bias whose magnitude increases in paralle
l to the dilution, is indicative of an sBC-dependent bias. The AXSYM FT4 as
say exhibited a significant sBC-dependent bias. This sBC-dependent bias is
likely to have been caused by the presence of significant amounts of T4 bin
ding proteins in the assay reagents.