S. Ropele et al., Fast multislice T-1 and T-1sat imaging using a phase acquisition of composite echoes (PACE) technique, MAGN RES M, 42(6), 1999, pp. 1089-1097
Citations number
28
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
An optimized multislice data acquisition scheme for phase acquisition of co
mposite echoes (PACE) imaging is presented. This new scheme uses a repetiti
on time equal to the mixing time and an appropriate phase cycling scheme, w
hich allows for more efficient exploitation of all composite echoes. These
modifications provide true multislice capability in parallel to a reduction
of the acquisition time by a factor >2 compared with the original PACE met
hod. Moreover, T-1 values can be obtained directly from phase images withou
t the use of a lookup table. Because of the symmetrical application of the
radio frequency pulses, this method is also well suited for T-1sat imaging.
Phantom studies showed a significantly better accuracy of the multislice f
ast PACE technique compared with a conventional two point method in multisl
ice acquisition mode, although precision was limited at high T-1 values. T-
1 and T-1sat measurements in brain tissue of eight healthy volunteers confi
rmed the stability of the fast PACE technique in a clinical setting. (C) 19
99 Wiley-Liss, Inc.