Low-T-c magnetometer used for active shielding in the frequency range dc-500 Hz in biomagnetic measurements

Citation
D. Platzek et H. Nowak, Low-T-c magnetometer used for active shielding in the frequency range dc-500 Hz in biomagnetic measurements, SUPERCOND S, 12(11), 1999, pp. 940-942
Citations number
7
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science
Journal title
SUPERCONDUCTOR SCIENCE & TECHNOLOGY
ISSN journal
09532048 → ACNP
Volume
12
Issue
11
Year of publication
1999
Pages
940 - 942
Database
ISI
SICI code
0953-2048(199911)12:11<940:LMUFAS>2.0.ZU;2-1
Abstract
Ultra low-noise de SQUIDS have been used to detect disturbances in biomagne tic measurements and to feed an active shielding system that compensates th ese disturbances. This work describes some properties of the SQUIDs used an d the principle of active shielding that has been improved for frequencies from de to 500 Hz. So far an additional attenuation of up to 50 dB for near -de and almost 20 dB for 50 Hz magnetic disturbances has been achieved.