Cg. Fu et al., Determination of oxalic acid in urine by co-electroosmotic capillary electrophoresis with amperometric detection, TALANTA, 50(5), 1999, pp. 953-958
Co-electroosmotic capillary electrophoresis with amperometric detection at
a Cobalt phthalocyanine (CoPC) modified carbon paste electrode was evaluate
d for the determination of oxalic acid in urine. The running buffer consist
ed of 10 mM phosphate (pH = 5.70) and 0.25 mM Cetyltrimethylammonium bromid
e. Under the optimum conditions, a detection limit of 0.12 mu M was achieve
d for oxalic acid. The response was linear between 0.5 and 1000 mu M with a
correlation coefficient of 0.9995. Applications of the method to real urin
e samples were described. (C) 1999 Elsevier Science B.V. All rights reserve
d.