Type of screw does not influence holding power in the femoral head - A cadaver study with shearing test

Citation
Jy. Jenny et al., Type of screw does not influence holding power in the femoral head - A cadaver study with shearing test, ACT ORTH SC, 70(5), 1999, pp. 435-438
Citations number
11
Categorie Soggetti
Ortopedics, Rehabilitation & Sport Medicine","da verificare
Journal title
ACTA ORTHOPAEDICA SCANDINAVICA
ISSN journal
00016470 → ACNP
Volume
70
Issue
5
Year of publication
1999
Pages
435 - 438
Database
ISI
SICI code
0001-6470(199910)70:5<435:TOSDNI>2.0.ZU;2-Z
Abstract
We studied the holding power during cut-out tests of 4 different screw fixa tion devices of the femoral head, correlated to head size and bone density, on 48 paired deep frozen proximal femurs of fresh gross specimens. We foun d no significant effect of the device on the holding power. There was a sig nificantly higher maximal load at failure in large and dense heads. It seem s then unnecessary to advocate more sophisticated but also more expensive d evices to fix femoral neck fractures.