Jc. Otteni et al., Anaesthetic machine: fresh gas delivery units. I. Mechanical units with rotameter block and calibrated vaporizers, ANN FR A R, 18(9), 1999, pp. 956-975
Objectives: To analyse the design, performance, failures, the checking befo
re use of mechanical fresh gas delivery units (FGDUs) equipped with bobbin
or ball-flowmeters delivering a continuous gas flow and calibrated vaporize
rs, marketed in France in 1999.
Data sources: Articles were obtained from a Medline(R) review (1960-1999; s
earch terms: anaesthetic machine, flowmeter, vaporizer), textbooks and pers
onal files; specific data were provided by manufacturers.
Study selection and data extraction: The articles were considered for perfo
rmance data, benefits and drawbacks, and characteristics, as well as the ri
sks carried by mechanical FGDUs.
Data synthesis: Seven anaesthetic machines out of 11 are equipped with mech
anical FGDUs, including rotameters delivering a gas mixture up to 30 L.min(
-1), calibrated vaporizers and an O-2-flush valve delivering at least 500 m
L O-2 per second (30 L.min(-1)). These units allow closed circuit anaesthes
is. They carry a risk for barotrauma as three out of them can deliver at th
e gas outlet of the FGDUs a gas mixture at a pressure reaching 3.5 bars and
four others at a pressure of 150-200 mmHg. They also carry a risk for hypo
xia, either from a preferential leak of oxygen at the corresponding rotamet
er or the O-2-flush valve, or from a leak of fresh gas mixture either in a
vaporizer or the selectatec manifold. The vaporizers carry a risk for vapor
delivery at a concentration differing notably from the value set on the co
ncentration dial. Therefore their accurracy must be checked periodically an
d the FGDUs checked for a leak after the addition to or the removal of a va
porizer from the selectatec manifold. The optimal technique for leak detect
ion is the negative pressure test. (C) 1999 Editions scientifiques et medic
ales Elsevier SAS.