Improving the shape and compliance characteristics of a high-volume, low-pressure cuff improves tracheal seal

Citation
Pj. Young et Mc. Blunt, Improving the shape and compliance characteristics of a high-volume, low-pressure cuff improves tracheal seal, BR J ANAEST, 83(6), 1999, pp. 887-889
Citations number
8
Categorie Soggetti
Aneshtesia & Intensive Care","Medical Research Diagnosis & Treatment
Journal title
BRITISH JOURNAL OF ANAESTHESIA
ISSN journal
00070912 → ACNP
Volume
83
Issue
6
Year of publication
1999
Pages
887 - 889
Database
ISI
SICI code
0007-0912(199912)83:6<887:ITSACC>2.0.ZU;2-9
Abstract
A prototype design of a compliant latex, high-volume, low-pressure cuffed t racheal tube cuff (CHVLP) was compared with the Mallinckrodt Hi-Lo, Sherida n preformed and Portex Profile high-volume, low-pressure (HVLP) cuffed trac heal tubes for leakage of dye placed above the cuff in a benchtop mechanica l ventilation model and in five isolated pig tracheas. There was no leakage in the ventilation model or in the pig tracheas with the prototype CHVLP. There was rapid leakage in the ventilation model and in all the pig trachea s for the Mallinckrodt Hi-Lo, the Sheridan preformed and the Porter Profile cuffs. This benchtop study suggests that improved HVLP cuff compliance cha racteristics may be beneficial in the prevention of leakage of fluid to the lungs known to occur with HVLP cuffs.