Infusion pump performance in an MR environment

Citation
Ej. Williams et al., Infusion pump performance in an MR environment, EUR J ANAES, 16(7), 1999, pp. 468-472
Citations number
3
Categorie Soggetti
Aneshtesia & Intensive Care
Journal title
EUROPEAN JOURNAL OF ANAESTHESIOLOGY
ISSN journal
02650215 → ACNP
Volume
16
Issue
7
Year of publication
1999
Pages
468 - 472
Database
ISI
SICI code
0265-0215(199907)16:7<468:IPPIAM>2.0.ZU;2-J
Abstract
Concerns about life support equipment accompanying the critically ill patie nt have to date made magnetic resonance imaging (MRI) studies of this patie nt group the exception. We present here a series of tests performed on an I VAC P3000 infusion pump to investigate its suitability for the magnetic res onance imaging environment. We investigate safety, pump performance and ima ge quality issues. The pump was housed at the end of the patient couch to p revent motion towards the scanner. Gravimetric tests found the pump to work within acceptable parameters at a static field of 10 mT. Image interferenc e issues were addressed.