Xd. Tian et al., Analytical performance of axial inductively coupled plasma time of flight mass spectrometry (ICP-TOFMS), J ANAL ATOM, 14(12), 1999, pp. 1807-1814
Analytical performance of an axial ICP-TOFMS, in terms of accuracy, precisi
on, resolution, signal stability, repeatability, reproducibility, precision
of isotope ratios and detection limits is reported. Cool plasma conditions
(low power and high central gas flow) allow the determination of K, Ca and
Fe at trace levels. Detection limits for 64 elements are reported with typ
ical values from 0.5-20 pg ml(-1) (3 s criterion). The long-term stability
over 4 h for the raw signal ranges between 1.0% RSD for Li-7 to 2.1% RSD fo
r Ca-40 operating under cool plasma conditions. Under normal plasma conditi
ons the corresponding values are 6.9% RSD for Pb-208 to 12.8% RSD for Co-59
. Accuracy for the determination of 16 different elements is assessed by me
asuring NIST 1643d (trace elements in water), yielding results in agreement
with certified values. Results for isotopic ratio determinations demonstra
te that a precision of 0.07-0.7% RSD is obtained within a short data acquis
ition period.