PHASE VALIDATION CRITERIA OF SIZE MEASUREMENTS FOR THE PHASE DOPPLER TECHNIQUE

Citation
Y. Hardalupas et Amkp. Taylor, PHASE VALIDATION CRITERIA OF SIZE MEASUREMENTS FOR THE PHASE DOPPLER TECHNIQUE, Experiments in fluids, 17(4), 1994, pp. 253-258
Citations number
19
Categorie Soggetti
Mechanics,"Engineering, Mechanical
Journal title
ISSN journal
07234864
Volume
17
Issue
4
Year of publication
1994
Pages
253 - 258
Database
ISI
SICI code
0723-4864(1994)17:4<253:PVCOSM>2.0.ZU;2-J
Abstract
A validation procedure of the size measurements for the phase Doppler anemometer is proposed for a two and three photodetector system and is shown experimentally to reduce erroneous measurements due to unwanted Doppler signals caused by surface reflection rather than refraction t hrough the particles. This problem known as 'trajectory ambiguity' is caused by the Gaussian distribution of the incident laser beams when t he particle diameter is similar to the diameter of the probe volume of the anemometer. The three photodetector system is shown to improve fu rther the sizing accuracy of a two photodetector system by rejecting m easurements from signals with phase ratio values lying outside the exp ected value for spherical scatterers and which would otherwise have be en validated by the two photodetector system.