Novel calibration method for filter radiometers

Citation
Ca. Schrama et H. Reijn, Novel calibration method for filter radiometers, METROLOGIA, 36(3), 1999, pp. 179-182
Citations number
12
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences","Instrumentation & Measurement
Journal title
METROLOGIA
ISSN journal
00261394 → ACNP
Volume
36
Issue
3
Year of publication
1999
Pages
179 - 182
Database
ISI
SICI code
0026-1394(1999)36:3<179:NCMFFR>2.0.ZU;2-6
Abstract
We present a new method for the calibration of inhomogeneous irradiance or illuminance meters. Using this method, a lux meter equipped with a mosaic f ilter was calibrated with an estimated uncertainty of 0.15 % (1 sigma). A p reliminary illuminance comparison has been performed between the new scale and a lamp-based scale maintained at the Nederlands Meetinstituut Van Swind en Laboratorium (NMi VSL). The observed difference is equal to the uncertai nty of this comparison, 0.4 %.