An amperometric glucose-sensing electrode based on the cathodic detection o
f oxygen consumption during the glucose oxidase (GOx) reaction was prepared
as follows. A platinum electrode was modified with polydimethylsiloxane (P
DMS), and a GOx-layer was placed on the PDMS layer. The permselectivity of
PMDS made it possible to monitor the oxygen consumption rate without interf
erence from the hydrogen peroxide. The concentration of glucose (0.01-1.2 m
M) could be determined from the decrease in the cathodic current at -0.1 V
vs. Ag/AgCl, without any interference from uric acid, acetaminophen and L-c
ysteine.