Copper determination after FI on-line sorbent preconcentration using 1-nitroso-2-naphthol as a complexing reagent

Citation
A. Ali et al., Copper determination after FI on-line sorbent preconcentration using 1-nitroso-2-naphthol as a complexing reagent, FRESEN J AN, 365(8), 1999, pp. 642-646
Citations number
23
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
FRESENIUS JOURNAL OF ANALYTICAL CHEMISTRY
ISSN journal
09370633 → ACNP
Volume
365
Issue
8
Year of publication
1999
Pages
642 - 646
Database
ISI
SICI code
0937-0633(199912)365:8<642:CDAFOS>2.0.ZU;2-O
Abstract
The suitability of 1-nitroso-2-naphthol as a complexing agent for on-line p reconcentration of copper using RP-C-18 material in a microcolumn with flow injection coupled with flame atomic absorption spectrometry (FI-FAAS) has been tested. Various parameters affecting complex formation, such as pH, sa mple flow rate, etc, and its elution into the nebulizer of FAAS were optimi zed. A 5 x 10(-3) mol/L reagent was on-line mixed with aqueous sample solut ion acidified with 0.1% (v/v) nitric acid (pH 3-4) and flowed through the m icrocolumn for 30 s. The adsorbed complexes in the microcolumn were eluted with ethanol in 10 a into the nebulizer of FAAS. A good precision (1.7% for 50 mu g/L copper, n = 12), high enrichment factor (19) with detection limi t (3 sigma) 2.0 mu g/L, and sample throughput (90 h(-1)) were obtained. The method was applied to certified reference materials seawater, mussel (biol ogical), NBS-362 and NBS-364 (special low alloy steel), for the determinati on of copper, and the results were in good agreement with the certified val ues.