Comparative measurements on a series of accelerators by the same vendor

Authors
Citation
Rj. Watts, Comparative measurements on a series of accelerators by the same vendor, MED PHYS, 26(12), 1999, pp. 2581-2585
Citations number
5
Categorie Soggetti
Radiology ,Nuclear Medicine & Imaging","Medical Research Diagnosis & Treatment
Journal title
MEDICAL PHYSICS
ISSN journal
00942405 → ACNP
Volume
26
Issue
12
Year of publication
1999
Pages
2581 - 2585
Database
ISI
SICI code
0094-2405(199912)26:12<2581:CMOASO>2.0.ZU;2-2
Abstract
A series of comparative measurements were conducted during a 12-month perio d on a series of linear accelerators made by the same vendor. These measure ments were conducted during acceptance and commissioning of six Varian mode l 2100C linear accelerators which had identical beam energies and beam modi fiers. Nominal accelerating potentials, scatter measurements, depth doses, beam profiles, wedge and electron beam descriptors are presented. The resul ts show highly similar beam parameters with observed variations being appro ximately equivalent to the precision of measurement. This similarity implie s that a standard data set can be used to validate new commissioning data. (C) 1999 American Association of Physicists in Medicine. [S0094-2405(99)001 12-1].