Active laser cavity particle counters: a fundamental problem and solution for airborne application

Citation
Jm. Rosen et Nt. Kjome, Active laser cavity particle counters: a fundamental problem and solution for airborne application, APPL OPTICS, 38(36), 1999, pp. 7321-7324
Citations number
10
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science","Optics & Acoustics
Journal title
APPLIED OPTICS
ISSN journal
00036935 → ACNP
Volume
38
Issue
36
Year of publication
1999
Pages
7321 - 7324
Database
ISI
SICI code
0003-6935(199912)38:36<7321:ALCPCA>2.0.ZU;2-4
Abstract
A previously unreported air-density-related detuning effect for active cavi ty aerosol optical particle counters is described, and a simple device for correcting the problem is presented. Airborne field tests with and without the correction device illustrate the effect and the good degree to which th e deficiency can be overcome. The effect is shown to have an important impa ct on instrument calibration. (C) 1999 Optical Society of America OCIS code s: 010.1080, 010.1100, 120.0120, 120.4820, 140.3460.