Feasibility of transmission x-ray microscopy at 4 keV with spatial resolutions below 150 nm

Citation
B. Kaulich et al., Feasibility of transmission x-ray microscopy at 4 keV with spatial resolutions below 150 nm, APPL PHYS L, 75(26), 1999, pp. 4061-4063
Citations number
15
Categorie Soggetti
Apllied Physucs/Condensed Matter/Materiales Science
Journal title
APPLIED PHYSICS LETTERS
ISSN journal
00036951 → ACNP
Volume
75
Issue
26
Year of publication
1999
Pages
4061 - 4063
Database
ISI
SICI code
0003-6951(199912)75:26<4061:FOTXMA>2.0.ZU;2-3
Abstract
The feasibility of a transmission x-ray microscope operating in the 3-7 keV photon energy range using phase zone plates (ZPs) has been tested. It has been demonstrated that, at a photon energy of 4 keV, structures smaller tha n 150 nm can be resolved with good contrast using exposure times in the sec ond to minute range. A large diameter gold ZP was used as condenser and a t ungsten ZP with an outermost zone width of 128 nm was used as a high spatia l resolution imaging objective lens. Images with a field size of 10 x 10 mu m(2) were acquired using a charge coupled device camera optically coupled to a phosphor screen. The corresponding pixel size in the object plane was 70 nm. (C) 1999 American Institute of Physics. [S0003-6951(99)03952-2].