Experimental determination of energy resolution and transmission characteristics of an electrostatic toroidal spectrometer adapted to a standard scanning electron microscope
D. Berger et al., Experimental determination of energy resolution and transmission characteristics of an electrostatic toroidal spectrometer adapted to a standard scanning electron microscope, J ELEC SPEC, 105(2-3), 1999, pp. 119-127
Citations number
27
Categorie Soggetti
Spectroscopy /Instrumentation/Analytical Sciences
Journal title
JOURNAL OF ELECTRON SPECTROSCOPY AND RELATED PHENOMENA
The main characteristics of sectorial toroidal energy analyzer for a new el
ectrostatic electron spectrometer adapted to a standard scanning electron m
icroscope are defined and determined experimentally. These transfer charact
eristics, i.e. intensity-energy response and transmission functions, energy
resolution and coupling constant, are needed for spectrometer calibration,
registration and energy correction of measured backscattered electron spec
tra and microtomographic analysis of multilayered structures. The spectrome
ter response to a monoenergetic primary electron beam and to a continuous e
nergy distribution is discussed. Detector response functions for energy ind
ependent and linear energy dependent detectors are considered. For aperture
slits of the spectrometer which allow reasonable electron intensities at t
he detector an energy resolution of about 2.5% is obtained. (C) 1999 Elsevi
er Science B.V. All rights reserved.