E. Bechgaard et al., HIGH-PERFORMANCE LIQUID-CHROMATOGRAPHIC ANALYSIS OF PEPTIDE-T IN RABBIT PLASMA WITH ONLINE COLUMN ENRICHMENT, Journal of chromatography B. Biomedical sciences and applications, 693(1), 1997, pp. 237-240
Citations number
8
Categorie Soggetti
Chemistry Analytical","Biochemical Research Methods
The present paper describes the development of a simple and sensitive
analytical method for quantification of Peptide T (PT) in rabbit plasm
a, using standard analytical equipment and on-line column enrichment,
without prior extraction, clean-up or derivatization. The analytical p
rocedure was found to be accurate, precise and linear. The accuracy wa
s 100% (range 97-103%) and the mean precision was 8% (range 3-14%) for
all (n=6) concentrations (0, 15, 50, 100 and 200 ng/ml). The total re
covery was found to be approximately 80%, and it was found to be depen
dent upon the injection rate onto the extraction column. The correlati
on between added and found concentrations was 0.9982, and the limit of
detection was estimated to be around 5 ng/ml. The method is therefore
found to be suitable for bioavailability studies, involving Peptide T
, in rabbits.