Ie. Panderi et M. Parissi-poulou, Simultaneous determination of benazepril hydrochloride and hydrochlorothiazide by micro-bore liquid chromatography, J PHARM B, 21(5), 1999, pp. 1017-1024
A micro-bore liquid chromatographic method was developed for the simultaneo
us determination of benazepril hydrochloride and hydrochlorothiazide in pha
rmaceutical dosage forms. The use of a BDS C-18 micro-bore analytical colum
n, results in substantial reduction in solvent consumption and increased se
nsitivity. The mobile phase consisted of a mixture of 0.025 M sodium dihydr
ogen phosphate (pH 4.8) and acetonitrile (55:45, v/v), pumped at a flow rat
e of 0.40 ml min(-1). Detection was set at 250 nm using an ultraviolet dete
ctor. Calibration graphs are linear (r better than 0.9991, n = 5), in conce
ntration range 5.0-20.0 mu g ml(-1) for benazepril hydrochloride and 6.2-25
.0 mu g ml(-1) for hydrochlorothiazide. The intra- and interday R.S.D. valu
es were < 1.25% (n = 5), while the relative percentage error (E-r) was < 0.
9% (n = 5). The detection limits attained according to IUPAC definition wer
e 0.88 and 0.58 mu g ml(-1) for benazepril hydrochloride and hydrochlorothi
azide, : respectively. The method was applied in the quality control of com
mercial tablets and content uniformity test and proved to be suitable for r
apid and reliable quality control. (C) 1999 Elsevier Science B.V. All right
s reserved.